Improved Imaging in low Energy Electron Microscopy and Photo Emission Electron Microscopy Using Medipix2 Pixel Detector

Date Issued
2010-01-01
Author(s)
Sikharulidze, I.
Van Gastel, R.
Schramm, S.
Abrahams, J. P.  
Poelsema, B.
Trom, R. M.
Van der Molen, S. J.
DOI
10.1142/9789814307529_0023
Abstract
The application of the Medipix2 hybrid pixel detector in Low Energy Electron Microscopy (LEEM) and Photo Emission Electron Microscopy (PEEM) led to an improvement of the recorded image quality compared to the original setup based on microchannel plate (MCP), phosphor screen and CCD. The measurements were performed on an Elmitec LEEM III instrument without energy filter using an Ir(111) sample with graphene islands grown on the surface. The Medipix2 images exhibited better resolution and higher contrast compared to the MCP data. The results suggest that Medipix2 has potential to become the detector of choice for LEEM/PEEM instruments.