Repository logo
Log In
  1. Home
  2. Unibas
  3. Publications
  4. Comparison of the signal-to-noise ratio at 1.5 and 2.0 T using a whole-body system
 
  • Details

Comparison of the signal-to-noise ratio at 1.5 and 2.0 T using a whole-body system

Date Issued
1989-01-01
Author(s)
Beckmann, N.
Muller, S.
Seelig, J.  
DOI
10.1002/mrm.1910090311
Abstract
Signal-to-noise ratios for 1H, 31P, and 13C signals at 1.5 and 2.0 T were compared using a whole-body system with a fast rampable magnet. Results indicate that SNR lies within the predicted values of linear and 7/4th power dependency with the applied magnetic field, for lossy and lossless samples, respectively.
University of Basel

edoc
Open Access Repository University of Basel

  • About edoc
  • About Open Access at the University of Basel
  • edoc Policy

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Privacy policy
  • End User Agreement