  4. The challenges of sodium measurements - indirect versus direct ion selective method
 
  • Details

The challenges of sodium measurements - indirect versus direct ion selective method

Date Issued
2019-01-01
Author(s)
Refardt, Julie  
Sailer, Clara O. Odilia
Chifu, Irina
Winzeler, Bettina  
Schnyder, Ingeborg
Fassnacht, Martin
Fenske, Wiebke Kristin
Christ-Crain, Mirjam  
DOI
10.1530/eje-19-0101
Abstract
Diagnosis and treatment of dysnatremia is challenging and further complicated by the pitfalls of different sodium measurement methods. Routinely used sodium measurements are the indirect (plasma/serum) and direct (whole blood) ion selective electrode (ISE) method, showing discrepant results especially in the setting of acute illness. Few clinicians are aware of differences between the methods in clinically stable patients or healthy volunteers.; Data of 140 patients and 91 healthy volunteers undergoing osmotic stimulation with hypertonic saline infusion were analyzed. Sodium levels were measured simultaneously by indirect and direct ISE method before and at different time points during osmotic stimulation up to a sodium threshold of ≥150 mmol/l. The primary outcome was the difference in sodium levels between the indirect and the direct ISE method.; 878 sodium measurements were analyzed. Mean (SD) sodium levels ranged from 141mmol/l (2.9) to 151mmol/ (2.1) by the indirect ISE compared to 140mmol/l (3) to 149mmol/l (2.8) by the direct ISE method. The interclass correlation coefficient between the two methods was 0.844 (95%-CI 0.823, 0.863). On average, measurements by the indirect ISE were 1.9mmol/l (95%-CI limits -3.2; 6.9) higher than by the direct ISE method (p<0.001). The tendency of the indirect ISE method resulting in higher levels increased with increasing sodium levels.; Intra-individual sodium levels differ significantly between the indirect and direct ISE method also in the absence of acute illness. It is therefore crucial to adhere to the same method in critical situations to avoid false decisions due to measurement differences.
University of Basel

edoc
Open Access Repository University of Basel

  • About edoc
  • About Open Access at the University of Basel
  • edoc Policy

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Privacy policy
  • End User Agreement