edoc

Validation of the PhysioQuant blood pressure measuring monitor in adults

Langewitz, Wolf and Tanner, Sabine. (2009) Validation of the PhysioQuant blood pressure measuring monitor in adults. Blood pressure monitoring, Vol. 14. pp. 220-221.

Full text not available from this repository.

Official URL: http://edoc.unibas.ch/dok/A6003917

Downloads: Statistics Overview

Abstract

A new automatic blood pressure (BP) measuring device PhysioQuant (also marketed under the brand name Ergoscan) was evaluated according to the International Protocol for Validation of Blood Pressure Measuring Devices in adults issued by the European Society of Hypertension. BP values measured by the PhysioQuant were compared with BP readings from two independent observers. Thirty-three patients (13 male, 20 female) provided systolic and diastolic BP readings in the normotensive, borderline hypertensive, and hypertensive range. Their age varied between 29 and 93 years, their arm circumference between 23.5 and 37 cm. The device showed a mean (+/-SD) deviation from observer measurements of -0.1 (+/-3.4) mmHg for systolic and +1.1 (+/-3.9) mmHg for diastolic BP. The accuracy of the device did not vary according to BP values or other patient characteristics. The device passed all phases of the protocol and can be recommended following the regulation rules of the European Society of Hypertension.
Faculties and Departments:03 Faculty of Medicine > Bereich Medizinische Fächer (Klinik) > Psychosomatik > Psychosomatik (Langewitz)
03 Faculty of Medicine > Departement Klinische Forschung > Bereich Medizinische Fächer (Klinik) > Psychosomatik > Psychosomatik (Langewitz)
UniBasel Contributors:Langewitz, Wolf Axel
Item Type:Article, refereed
Article Subtype:Research Article
Publisher:Rapid Science Publ.
Note:Publication type according to Uni Basel Research Database: Journal article
Related URLs:
Identification Number:
Last Modified:26 Apr 2013 07:02
Deposited On:26 Apr 2013 06:52

Repository Staff Only: item control page